Declaration of Conformity UKCA CE for the OP5705XG and the OP5705-IO
View/Download the Declaration of Conformity CE-UKCA in PDF at the bottom of the page.
Manufacturer Name/Address:
OPAL-RT Technologies Inc.
1751 Richardson suite 1060
Montreal, QC H3K 1G6 Canada
www.opal-rt.com
This declaration of conformity is issued under the sole responsibility of: OPAL-RT Technologies Inc.
Product Model/Product
OP5705XG, OP5705-IO with any of the following options:
OP5330-3
OP5340
OP5342
OP5353
OP5360-3
OP5367-1
OP5367-3
OP5367-5
OP5368
OP5369
OP3411 (CAN kit)
OP3811 / OP3812 (Oregano kit)
OP5455 (Copper PCIe kit)
PXH832 (Dolphin optical PCIe kit)
I350T4V2BLK (Ethernet)
Product Description
Simulator with Digital and Analog Input and Output cards intended for indoor use only.
Ratings: 100-240 VAC, 50-60 Hz, 8.0 A - 4.0 A, 600W max, pollution degree 2
Product Category
Electrical equipment for measurement, control, and laboratory use.
CE Conformity with the Relevant Community Harmonization Legislation
2014/30/EU – Electromagnetic Compatibility Directive
2014/35/EU – Low Voltage Directive
2011/65/EU & (EU) 2015/863 – Restriction of Hazardous Substances (RoHS) in Electrical and Electronic Equipment directive
2012/19/EU – Waste from Electrical and Electronic Equipment (WEEE) Directive
UKCA Conformity with the Relevant Harmonization Legislation
Electromagnetic Compatibility Regulations 2016
Electrical Equipment (Safety) Regulations 2016
The Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment Regulations 2012
References to the standards in relation to which conformity is declared | ||
---|---|---|
Category | Standards | Test Specifications |
Emissions: | EN61326-1 (2021) / EN61326-1 (2013) | |
EN55011 (2016) A1(2017) | Conducted Emissions: Group 1 - class A 150kHz-30MHz | |
Immunity: | IEC61000-4-2 (2008) | Contact: ±4kV Air: ±2kV, ±4kV, ±8kV |
IEC61000-4-3 (2020) | 80MHz-1000MHz: 10V/m | |
IEC61000-4-4 (2012) | Power: ±2kV / 5kHz & 100kHz | |
IEC61000-4-5 (2014) A1 (2017) | Power: ±2kV L-PE / ±1kV L-L | |
IEC61000-4-6 (2013) | Power: 3V | |
IEC61000-4-8 (2009) | Continuous Field: 30A/m – 50Hz & 60Hz | |
IEC61000-4-11 (2020) | Voltage dips: Short interruptions: | |
Safety: | EN 61010-1 / IEC 61010-1 :2010, IEC 61010-1 :2010/AMD1 :2016 | Test procedure: CB Scheme |