Known Issues 1.16.0

  • EFSNITT-788: When the Reset Models Channel is asserted, the models are reset and placed into a paused state, but some outputs of the models retain their last valid value (instead of going back to 0).

  • PEA-133: For frequencies below 1kHz, the PWM Generators produce signals with a frequency error. The observed error is approximately 0.02% for a 100Hz signal, 0.3% for a 10Hz signal, and 2.3% for a 1Hz signal.

  • PEATT-1081: In VeriStand 2024 Q3, error 1003 occurs when a circuit model is loaded. This is due to Known Issue 2798144 in version 2024 Q3 of the LabVIEW Runtime Engine. To resolve this issue, install Patch 1 for the LabVIEW Runtime Engine 2024 Q3:

  1. Close all NI software, including NI VeriStand.

  2. Open the NI Package Manager.

  3. Under the INSTALLED tab, locate the LabVIEW Runtime (64-bit) and confirm its version. If the version shown is ‘2024 Q3' or ‘24.3.0.49385-0+f233,’ then the patch is not installed.

  4. Navigate to the UPDATES tab and select LabVIEW Runtime (64-bit) version ‘2024 Q3 Patch 1’ or ‘24.3.1.4915-0+f2’.

  5. Click Update and reboot the machine when prompted.

 